Turbidity metr Turb® 355
Turbidity meter Turb® 355, Type Turb® 355 , Description with tungsten lamp
Zařazeno v kategoriích
Turbidimetry
Popis
Turbidity meter Turb® 355
Battery-operated portable turbidity meter with infrared LED for nephelometric measurements as per ISO 7027/DIN EN 27027 (EN ISO 7027) or as a white light model with tungsten lamp as per US EPA. It is handy, light and simple to operate and is suitable, primarily for applications to monitor > 1 NTU. With smaller samples of 15 ml, it is also very suited for applications in fermenters and cell cultures at good precision.
Scope of delivery: Turbidity meter in a handy small case with quick instructions, calibration standards 0.02-10.0-1000 NTU, empty cuvettes and batteries.
Specifications
Measuring range: NTU 0 to 1100
FNU 0 to 1100
Resolution: 0.01 NTU in the range 1 to 9.99
0.1 NTU in the range 10.0 to 99.9
1 NTU in the range 100 to 1000
Accuracy: 0 to 500 NTU/FNU: ± 0.1 NTU/FNU or ± 2 % of measured value
500 to 1100 NTU/FNU: ± 3 % of measured value
Reproducibility: 0.05 NTU or ± 1% of measured value
Cell size: 25 x 45 mm, 15 ml sample volume
Power supply: 4 x AAA batteries for approx. 1500 measurements
Type Turb® 355
Description with tungsten lamp
Battery-operated portable turbidity meter with infrared LED for nephelometric measurements as per ISO 7027/DIN EN 27027 (EN ISO 7027) or as a white light model with tungsten lamp as per US EPA. It is handy, light and simple to operate and is suitable, primarily for applications to monitor > 1 NTU. With smaller samples of 15 ml, it is also very suited for applications in fermenters and cell cultures at good precision.
Scope of delivery: Turbidity meter in a handy small case with quick instructions, calibration standards 0.02-10.0-1000 NTU, empty cuvettes and batteries.
Specifications
Measuring range: NTU 0 to 1100
FNU 0 to 1100
Resolution: 0.01 NTU in the range 1 to 9.99
0.1 NTU in the range 10.0 to 99.9
1 NTU in the range 100 to 1000
Accuracy: 0 to 500 NTU/FNU: ± 0.1 NTU/FNU or ± 2 % of measured value
500 to 1100 NTU/FNU: ± 3 % of measured value
Reproducibility: 0.05 NTU or ± 1% of measured value
Cell size: 25 x 45 mm, 15 ml sample volume
Power supply: 4 x AAA batteries for approx. 1500 measurements
Type Turb® 355
Description with tungsten lamp
Varianty produktu
Pro zobrazení cen na míru se přihlašte do klientské sekce.
Kód produktu
Kód produktu
Kód BDL 6902303 Kód výrobce 600311


Kód BDL 6236840 Kód výrobce
600312
Turbidity meter Turb® 355, Type Turb® 355 , Description with tungsten lamp
600312


Máte otázky? Kontaktujte nás
Poslat